02+ Thermocouple Calibrator
-
Simulation Specification
Output
Range
Output Range
Resolution
Accuracy
Illustration
DCV
100mV
-10.00~110.00mV
0.01mV
0.05%+30uV
Max. output current ±2mA
1V
-0.1000~1.1000V
0.1mV
0.05%+0.3mV
Thermocouple
R
-40~1760°C
1°C
0.05%+3 (≤100°C)
0.05%+2 (>100°C)
Employs ITS-90 temperature standard
The accuracy does not include the error of interior temperature compensation sensor
The accuracy does not include the impact of interior thermoelectric force
S
-20~1760°C
1°C
B
400~1800°C
1°C
0.05%+3 (≤ 600°C)
0.05%+2 (>600°C)
E
-200.0~1000.0°C
0.1°C
0.05%+20 (≤-100°C)
0.05%+10 (> -100°C)
K
-200.0~1370°C
0.1°C
J
-200.0~1200.0°C
0.1°C
T
-200.0~400.0°C
0.1°C
N
-200.0~1300.0°C
0.1°C
Measurements Specification
Input
Range
Output range
Resolution
Accuracy
Illustration
DCmV
100mV
-10.00~110.00mV
0.01mV
0.05%+3
Input resistance: 1MΩ
Thermocouple
R
-40°C~1760°C
1°C
0.05%+3(≤ 100°C)
0.05%+2 (>100°C)
Input resistance : 1MΩ
Employs ITS-90 temperature standard
The accuracy does not include the error of interior temperature compensation sensor
The accuracy does not include the impact of interior thermoelectric force
S
-200°C~1760°C
B
400°C~1800°C
0.05%+3(≤ 600°C)
0.05%+2 (> 600°C)
K
-200.0°C~1370.0°C
0.1°C
0.05%+20(≤ -100°C)
0.05%+10 (> -100°C)
E
-200.0°C~1000.0°C
J
-200.0°C~1200.0°C
T
-200.0°C~400.0°C
N
-200.0°C~1300.0°C
General Information
Power Supply 2* 1.5V batteries
Product Size 180(L)×90(W)×47(H)mm
Product Weight 500g
Standard Accessories User’s Manual, industrial testing lead (clips for probe)
Maximum Allowed Voltage :30V(within terminals or between terminal and earth ground)
Operation Condition :0~50°C / ≤ 80%RH
Storage Condition :-10~55°C / ≤ 90%RH
Temperature coefficient :0.1×(dedicated accuracy)%/°C(5~18°C, 28~40°C)
*Measures and simulates 8 thermocouple types and millivolts
*Temperature Unit ℃/℉Selectable
*High accuracy up to: 0.05%
* Auto range
*Manual and automatic cold junction compensation available
*Key-operation , easy to use